Abstract
This is a descriptive study designed to assess the predictive value of intraocular pressure (IOP) measurement in GPs' offices in an urban healthcare site using Tono-pen XL. A total of 2044 patients, aged > or =40 years, were enrolled by consecutive sampling from patients visiting the GP. Those participants who had IOP > or =21 mmHg were referred to the ophthalmologist. Of the 226 then tested, ocular hypertension was confirmed in 100 participants (4.89%, 95% CI [confidence interval] = 3.93 to 5.85%). Predictive value was 44.2%. These results suggest the validity of using Tono-pen XL in the GP's office to detect ocular hypertension.
Journal Title
Journal ISSN
Volume Title
Publisher
Royal College of General Practitioners
DOI
Date
Description
Keywords
Citation
Beneyto P, Barajas MA, Garcia-de-Blas F, Del Cura I, Sanz T, Vello R, Salvador C. Predictive value of tonometry with Tono-pen XL in primary care. Br J Gen Pract. 2007 Aug;57(541):653-4