Slice-specific B1+ shimming improves the repeatability of multishot DWI at 7T

dc.contributor.authorDing, Belinda
dc.contributor.authorWilliams, Sydney Nicole
dc.contributor.authorDragonu, Iulius
dc.contributor.authorLiebig, Patrick
dc.contributor.authorAllwood-Spiers, Sarah
dc.contributor.authorMcElhinney, Paul
dc.contributor.authorGunamony, Shajan
dc.contributor.authorFullerton, Natasha
dc.contributor.authorPorter, David Andrew
dc.date.accessioned2025-01-28T16:04:33Z
dc.date.available2025-01-28T16:04:33Z
dc.date.issued2024-08-01
dc.description.abstractPurpose: Compared with lower field strengths, DWI at 7 T faces the combined challenges of increased distortion and blurring due to B0 inhomogeneity, and increased signal dropouts due to B1+ inhomogeneity. This study addresses the B1+ limitations using slice-specific static parallel transmission (pTx) in a multi-shot, readout-segmented EPI diffusion imaging sequence. Methods: DWI was performed in 7 healthy subjects using MRI at 7 T and readout-segmented EPI. Data were acquired with non-pTx circular-polarized (CP) pulses (CP-DWI) and static pTx pulses (pTx-DWI) using slice-specific B1+ shim coefficients. Each volunteer underwent two scan sessions on the same day, with two runs of each sequence in the first session and one run in the second. The sequences were evaluated by assessing image quality, flip-angle homogeneity, and intrasession and intersession repeatability in ADC estimates. Results: pTx-DWI significantly reduced signal voids compared with CP-DWI, particularly in inferior brain regions. The use of pTx also improved RF uniformity and symmetry across the brain. These effects translated into improved intrasession and intersession repeatability for pTx-DWI. Additionally, re-optimizing the pTx pulse between repeat scans did not have a negative effect on ADC repeatability. Conclusion: The study demonstrates that pTx provides a reproducible image-quality increase in multishot DWI at 7 T. The benefits of pTx also extend to quantitative ADC estimation with regard to the improvement in intrasession and intersession repeatability. Overall, the combination of multishot imaging and pTx can support the development of reliable, high-resolution DWI for clinical studies at 7 T.
dc.identifier.citationDing B, Williams SN, Dragonu I, et al. Slice-specific B1+ shimming improves the repeatability of multishot DWI at 7 T. Magn Reson Med. 2024; 92: 2560-2570. doi: 10.1002/mrm.30208
dc.identifier.doihttps://doi.org/10.1002/mrm.30208
dc.identifier.issn1522-2594 (online)
dc.identifier.issn0740-3194 (print)
dc.identifier.urihttps://hdl.handle.net/10115/66217
dc.language.isoen_US
dc.publisherWiley
dc.rightsAttribution-NonCommercial-NoDerivatives 4.0 Internationalen
dc.rights.accessRightsinfo:eu-repo/semantics/openAccess
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/4.0/
dc.subjectdiffusion imaging
dc.subjectparallel transmission
dc.subjectrepeatability
dc.subjectRF shimming
dc.subjectultra-high field
dc.titleSlice-specific B1+ shimming improves the repeatability of multishot DWI at 7T
dc.typeArticle

Archivos

Bloque original

Mostrando 1 - 1 de 1
Cargando...
Miniatura
Nombre:
Ding_SliceSpecificB1Shimming_MRM_2024.pdf
Tamaño:
3.18 MB
Formato:
Adobe Portable Document Format