Influence of the biomechanical evaluation of rupture using two shapes of same intramedullary implant after proximal interphalangeal joint arthrodesis to correct the claw/hammer pathology: A finite element study

dc.contributor.authorBayod-López, Javier
dc.contributor.authorBecerro-de-Bengoa-Vallejo, Ricardo
dc.contributor.authorPrados-Frutos, Juan Carlos
dc.contributor.authorLosa-Iglesias, Marta
dc.contributor.authorLópez-López, Daniel
dc.contributor.authorPrados-Privado, María
dc.date.accessioned2024-11-04T08:38:17Z
dc.date.available2024-11-04T08:38:17Z
dc.date.issued2024
dc.description.abstractWe used finite element analysis to study the mechanical stress distribution of a new intramedullary implant used for proximal interphalangeal joint (PIPJ) arthrodesis (PIPJA) to surgically correct the claw-hammer toe deformity that affects 20% of the population. After geometric reconstruction of the foot skeleton from claw toe images of a 36-year-old male patient, two implants were positioned, in the virtual model, one neutral implant (NI) and another one 10° angled (10°AI) within the PIPJ of the second through fourth HT during the toe-off phase of gait and results were compared to those derived for the non-surgical foot (NSF). A PIPJA was performed on the second toe using a NI reduced tensile stress at the proximal phalanx (PP) (45.83 MPa) compared to the NSF (59.44 MPa; p < 0.001). When using the 10°AI, the tensile stress was much higher at PP and middle phalanges (MP) of the same toe, measuring 147.58 and 160.58 MPa, respectively, versus 59.44 and 74.95 MPa at corresponding joints in the NSF (all p < 0.001). Similar results were found for compressive stresses. The NI reduced compressive stress at the second PP (-65.12 MPa) compared to the NSF (-113.23 MPa) and the 10°AI (-142 MPa) (all p < 0.001). The von Mises stresses within the implant were also significantly lower when using NI versus 10°AI (p < 0.001). Therefore, we do not recommend performing a PIPJA using the 10°AI due to the increase in stress concentration primarily at the second PP and MP, which could promote implant breakage. Keywords: foot; foot deformities; foot diseases; musculoskeletal diseases.es
dc.identifier.citationBayod-López J, Becerro-de-Bengoa-Vallejo R, Prados-Frutos JC, Losa-Iglesias M, López-López D, Prados-Privado M. Influence of the biomechanical evaluation of rupture using two shapes of same intramedullary implant after proximal interphalangeal joint arthrodesis to correct the claw/hammer pathology: A finite element study. Int Wound J. 2024 Aug;21(8):e70014. doi: 10.1111/iwj.70014. PMID: 39107920; PMCID: PMC11303262.es
dc.identifier.doi10.1111/iwj.70014es
dc.identifier.urihttps://hdl.handle.net/10115/41035
dc.language.isoenges
dc.rightsAttribution-NonCommercial-NoDerivatives 4.0 International*
dc.rights.accessRightsinfo:eu-repo/semantics/openAccesses
dc.rights.urihttps://creativecommons.org/licenses/by-nc-nd/4.0/*
dc.titleInfluence of the biomechanical evaluation of rupture using two shapes of same intramedullary implant after proximal interphalangeal joint arthrodesis to correct the claw/hammer pathology: A finite element studyes
dc.typeinfo:eu-repo/semantics/articlees

Archivos

Bloque original

Mostrando 1 - 1 de 1
Cargando...
Miniatura
Nombre:
International Wound Journal - 2024 - Bayod‐López - Influence of the biomechanical evaluation of rupture using two shapes of.pdf
Tamaño:
6.71 MB
Formato:
Adobe Portable Document Format
Descripción:

Bloque de licencias

Mostrando 1 - 1 de 1
No hay miniatura disponible
Nombre:
license.txt
Tamaño:
2.67 KB
Formato:
Item-specific license agreed upon to submission
Descripción: